Click on the scale image to view the manual. Can't find what you're looking for? Call us at 1 (800) 815-6615 or email us at homprocs@homscales.com. Find the personal care product manual that you need at ManualsOnline. Health O Meter Scale BFM582. Health O Meter User Manual BODY FAT & HYDRATION MONITORING. HDM076DQ-60 - Health o meter® Weight Tracking Scale Manual; HDR900-01 - Health o meter® Digital Scale - White Manual; HDM166DQ-60 - Health o meter® Progress Bar Scale Manual; HDL155-01 - Health o meter® Digital Scale - White Manual; HDM169 - Health o meter® Weight Tracking Scale Manual; HDL830-63 - Health o meter® Digital Scale Manual. This Health o meter® Professional product is warranted from date of purchase against defects of materials or in workmanship for a period of two (2) years. If product fails to function properly, return the product, freight prepaid and properly packed to Pelstar, LLC (see “To Get Warranty Service”, below, for instructions). Repairs should only be performed by authorized Health o meter® Professional service personnel. To prevent patient injury, the patient must be attended throughout the entire weighing event. SPECIFICATIONS FOR THIS SCALE. The Model 402LB physician beam scale employs a proven and rugged mechanical design to measure patient weight.

  1. Health-o-meter Professional Scale Manual
  2. Health O Meter Hy 200 User Manual Free
  3. Health-o-meter Professional Scale User Ma…
  4. Health O Meter Hy 200 User Manual Online
Advertising

6

TROUBLESHOOTING

498KL

Refer to the following instructions to check and correct any failure before contacting service personnel.

SYMPTOM

POSSIBLE CAUSE

CORRECTIVE ACTION

Scale does not turn on

Meter

1. Battery problem

2. Bad Power Supply or Not Plugged In

3. Electrical Outlet not functioning

1. Check / Replace the batteries

2. Check / Replace Power Supply Adapter

3. Use a different outlet

“LobAt”

shows in display

The batteries are low

Replace batteries according to instructions

Questionable weight or

the scale does not zero

1. External object interfering with the scale

2. The display did not show “0.0” before

weighing

3. Scale is not placed on a level surface

4. Scale is out of calibration

1. Remove interfering object from the scale

2. Ask the patient to step off the scale. Zero the

scale and begin weighing process again.

3. Place scale on level surface

4. Check weight with know weight value

“OVER”

shows in display

Overload. The load on the scale exceeds scale

capacity (500 lbs / 220 kg)

Remove weight, turn scale off, turn scale on again

and try again

“UNDER”

shows in display

Negative weight, indicating a tare was present.

Load on the scale is below initial zero point

Clear scale of any weight and re-zero scale by

pressing ZERO button


CALIBRATION

This scale has been factory calibrated and does not require calibration prior to use.

The calibration procedure may be performed using a known weight in either pounds or kilograms. Simply set the scale for the desired

unit of measure using the lb/kg key and start the calibration procedure. The calibration is performed using lb or kg, according to the unit

of measure used upon entering into the calibration path, meaning the scale will default to either lb or kg, if that was the setting left on

the scale for weighing. Note: Calibration weight must be no less than 100 lb (40 kg) and no more than 500 lb (200 kg).

ACTION

DISPLAY SHOWS

Press and hold the LB/KG and ZERO buttons at the same time, then press the ON/OFF
button to turn on the scale. “CAL” followed by a number indicates you can calibrate.

“CAL” then

“XXXXX”

Press the ZERO button. “C-100” (for lb mode) or “C-40” (for kg mode) will flash on the

display and then the display will return to “0”.

“C-100” or “C-40”

Health O Meter Hny200 User Manual

Health-o-meter Professional Scale Manual

Place 100 lbs or 40 kg on the scale and when calibration number stabilizes, press the ZERO

The firmware and configuration tool can also change. Please see this commit:Installation Manual and User GuideThese can be found here:Please always use the version from above link, as we are continuosly improving the guide. Instead, please register to our community forum at and use appropriate thread there:End-user license and copyright informationPlease read the EULA.md file. User manual. Support and bug reportingPlease do not send support requests regarding the beta version to the Granite Devices support emails.

button. “C-200” (for lb mode) or “C-80” (for kg mode) will flash on the display and then the

display will return to “0”.

“C-200” or “C-80”

Place 200 lbs or 80 kg on the scale and when calibration number stabilizes, press the ZERO

button. “C-300” or “C-120” will appear on the display and then the display will return to “0”.

“C-300” or “C-120”

Place 300 lbs or 120 kg on the scale and when calibration number stabilizes, press the ZERO

button. “C-500” or “C-200” will appear on the display and then the display will return to “0”.

“C-500” or “C-200”

Place 500 lbs or 200 kg on the scale and when calibration number stabilizes, press the ZERO